PN B10093AE
2-9
Operation
Manual Operation
2
•
If you enter a value out of the achievable
2
t
range for that rotor, a message prompts you
to enter a different value.
4
Press the next function key or press
ENTER
, then
START
.
•
The centrifuge begins calculating and displaying the accumulated centrifugal effect.
•
Deceleration begins when the set
2
t
value is reached. During deceleration, the red
indicator light on the
STOP
key flashes.
•
The
2
t
continues to accumulate until the rotor stops spinning, so set and actual
2
t
values may be different at the end of the run.
Entering a Continuous (HOLD) Run
1
Press
TIME
.
The
HH:MM
,
W2T
, and
HOLD
soft keys appear.
2
Press the
HOLD
soft key to select the hold mode.
HOLD
appears in the
Time:
field.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
ROTOR ID
SPEED
RPM/RCF
TIME
w2t
°C
TEMP
Rotor : JA-30.50
Speed : 26 000 RPM = 81 800 x G
Time : 1.05E03 w2t
Temp : 25°C Max Temp : 29°C
HH:MM
W2T
HOLD
Accel : MAX Decel : MAX
EXP
03
rad2/sec
ROTOR ID
SPEED
RPM/RCF
TIME
HR:MIN
°C
TEMP
Rotor : JA-30.50
Speed : 26 000 RPM = 81 800 x G
Time : 00:00
Temp : 25°C Max Temp : 29°C
HH:MM
W2T
HOLD
HOLD
Accel : MAX Decel : MAX
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......