Operation
Entering Run Parameters
Figure 3.1
Acceleration and Deceleration Settings Graph
Setting the Acceleration Rate
NOTE
The default setting for both acceleration and deceleration is maximum (
MAX
). If the
MAX
indicator
lights are already lit, no input is required. To choose a different setting, follow the steps below.
Table 3.2
Deceleration Settings
Maximum Deceleration
Slow Deceleration
Off
Full brake is used from set
speed to near 0 rpm, to
bring the rotor to a stop in
the shortest possible time.
Reduced brake is used
during the last few rpm
until 0 is reached, to
minimize disturbance of the
sample.
Reduced torque is used from set speed to
500 rpm, causing deceleration to 500 rpm to
take twice as long as maximum deceleration.
(The exact rate depends on the mass of the
rotor in use. Larger rotors will take longer.)
Deceleration time from 500 to 0 rpm is
approximately 2 minutes (larger rotors take
longer).
No brake is used. Rotor
coasts to 0 rpm. This
process can take up to
1 hour, depending on the
rotor in use and the speed
at the start of
deceleration.
Application: pelleting
Application: gradients, pelleting
Application: gradients
1.
Rotor Speed (RPM)
2.
Time
3.
Max Accel
4.
Slow Accel
5.
Max Decel
6.
Run Speed
7.
Off (no brake)
8.
Slow Decel
0
500
1
2
3
4
5
6
7
8
Summary of Contents for Avanti J-26 XP Series
Page 6: ...J326XP IM 5AB vi Safety Notice Chemical and Biological Safety ...
Page 12: ...xii Tables ...
Page 16: ...J326XP IM 5AB xvi Introduction Recycling Label ...
Page 40: ...J326XP IM 5AB 2 8 Description Controls and Indicators ...
Page 54: ...J326XP IM 5AB 3 14 Operation Unloading the Centrifuge ...
Page 74: ...J326XP IM 5AB Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 76: ... 2012 Beckman Coulter Inc All Rights Reserved www beckmancoulter com ...