J326XP-IM-5AB
1-3
Specifications and Preinstallation Requirements
Available Rotors
1
Available Rotors
Refer to the applicable rotor manual for complete rotor descriptions.
Electrical Requirements
•
200/208/240-V, single-phase instrument
—
180–264 VAC, 30 A, 50/60 Hz
•
230-V, single-phase instrument
—
180–264 VAC, 30 A, 50 Hz
•
220/380-V plus neutral, three-phase
a
instrument
—
313–457 VAC plus neutral, 16 A, 50 Hz
Electrical Supply
Class I
Maximum Heat
dissipation into Room
Under Steady-State
Conditions
2.0 kW (6900 BTu/h)
Humidity restrictions
<95% (noncondensing)
Noise Level 0.91 m (3 ft)
in front of instrument at
26 000 rpm
57 dBa
Installation
(overvoltage) Category
II
Pollution Degree
2
b
a. Unbalanced three-phase. Split for single-phase operation internally.
b. Normally, only nonconductive pollution occurs; occasionally, however, a temporary conductivity caused by condensation
must be expected.
Specification
Description
Rotor Profile
Description
Rotor
Code
Max
RPM
a
Max
RCF
(
g
)
Max Capacity
Rotor
Manual
Number
JA-30.50 Ti
b
Fixed Angle, 34°
(8 place)
r
max
= 108 mm
30.50
26 000
c
81 800
8
50 mL
J-TB-070
JA-25.50 Ti
Fixed Angle, 34°
(8 place)
d
r
max
= 108 mm
25.50
25 000
75 600
8
50 mL
J-TB-056
Summary of Contents for Avanti J-26 XP Series
Page 6: ...J326XP IM 5AB vi Safety Notice Chemical and Biological Safety ...
Page 12: ...xii Tables ...
Page 16: ...J326XP IM 5AB xvi Introduction Recycling Label ...
Page 40: ...J326XP IM 5AB 2 8 Description Controls and Indicators ...
Page 54: ...J326XP IM 5AB 3 14 Operation Unloading the Centrifuge ...
Page 74: ...J326XP IM 5AB Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 76: ... 2012 Beckman Coulter Inc All Rights Reserved www beckmancoulter com ...