7-8
Analyzer Modes
7.5 Perform an Emergency Stop
7.4.2 Return to
Standby
Mode
1
Select
Stop/Standby
in Stop mode. The Warmup/Standby window
opens with the message “Reset Analyzer and Transfer to Standby
mode?”.
2
Select
OK
. The system reset operation is performed. After the reset
operation completes, the system goes to Standby mode or Warm up
mode.
3
Perform a W1.
For details on performing a W1, refer to 8.8.13 Perform a W1 Procedure in the
AU5800 User’s Guide, chapter 8.
7.5 Perform an Emergency Stop
Select the
EM STOP
button (red button on the rack feeder unit) to immediately
turn off power to the analyzer and ISE unit(s). All analysis immediately stops.
CAUTION
•
When the EM STOP button is pressed, any data that is not
complete is lost and must be repeated.
•
Reagents remain in the cuvettes after performing an
Emergency Stop. Perform a W1 to remove reagents before
performing an End Process or resuming analysis. For
details on performing the W1, refer to 8.8.13 Perform a W1
Procedure in the
AU5800 User's Guide
, chapter 8.