B00418AA
7-3
Analyzer Modes
7.1 Analyzer Modes
7
CAUTION
If a Stop or Emergency Stop occurs, sample can remain in the
sample probe, and reagents can remain in the cuvettes. Perform a
W1 to clean the sample probe and cuvettes.
For details on how to perform a W1, refer to 8.8.13 Perform a W1 Procedure in
the AU5800 User’s Guide, chapter 8.
Measure 1
Measure 1 occurs when the
Start
button is selected. Racks
are present on the rack feeder unit, and moving to the rack
transport unit.
Measure 2
Measure 2 occurs when there are no more racks on the rack
feeder unit. To start additional racks, select the
Start
button.
Stop
Stop
mode occurs when there is a system error, or when the
operator selects the
Stop/Standby
button. The analyzer can
not be started from
Stop
mode. To return to
Standby
mode,
select the
Stop/Standby
button. The mode displays as
Reset
while the hardware is initializing, then it goes to
Standby
. All
tests in progress must be repeated.
Pause
Pause
mode occurs when there is a system error, or when
the operator selects the
Pause
button. Analysis can be re-
started from
Pause
by selecting
Start
button. All tests in
progress will be completed.
Mode
Contents
B00418AA_AU5800 QRG_EN.book Page 3 Monday, December 13, 2010 9:59 AM