B00418AA
6-47
As Needed Tasks
6.16 Disable a Test
6
7
The Perform window opens when the operation is complete.
Figure 6.1
8
Select
OK
. Remove the external media.
6.16 Disable a Test
Specific tests can be selected to stop analysis (disable) even when the test has a
requisition. Examples of when it may be useful to disable a test are if the
calibration failed for that test, or QC fails and samples are in process.
TIP
A test can be disabled or enabled during a Measure mode. Analysis of the
test(s) will stop or re-start after selection in
Disable (F7)
.
Tests are not disabled for reagent blank, calibration, and QC samples.
Tests that are disabled display and print with a “/” error flag, indicating the
test was requisitioned but not performed.
Selections in the Disable window are effective until a new index is made,
or an End Process is performed.