B00418AA
6-9
As Needed Tasks
6.2 Create a Profile
6
13
Run the test to verify the programming of the new test(s).
•
Load the reagent and any required cleaning solution on the
analyzer.
•
Perform a reagent volume check.
•
Verify that calibration is requisitioned for the new test(s).
•
Requisition QC on the new test(s) if the test(s) was not added to
the default QC requisition.
•
Perform a reagent blank, calibration, and QC on the new test(s).
•
Review the printout and verify the reagent blank, calibration, and
QC data.
6.2 Create a Profile
A profile is a group of tests that are usually requested at the same time. Using a
profile reduces the number of key strokes needed, as a single profile is selected
instead of multiple tests. A maximum of 100 profiles (Number 0 to Number
99) can be registered for samples, reagent blank, calibration, and QC. A
maximum of 120 tests can be programmed in a profile. The number of tests
that can be programmed in a profile is limited by the number of analyzer units,
sample blank tests, LIH, and sample type.
Each profile is assigned a profile name.
TIP
Profile 0 is the Default profile in the
Sample
tab. Profile 0 is automatically
performed when:
•
There is a barcode read error.
•
There is no requisition found for a sample.
•
There are online errors.
TIP
Tests that are grayed out cannot be selected.
ISE tests can be selected only when the sample type is “Serum” or “Urine”.