92 - Documenting Calibrator
Figure 83: Button for zeroing a pressure module (lower right corner)
Before you tap the
Start
button, ensure that the signals are "live", i.e.
measurements and the required connections are as they should. If unsure,
use the connection diagrams found in the Instrument Overview Window. It
opens from the
Info
button.
If gauge pressure modules are used in the calibration, remember to zero
them before starting the calibration.
Tap the
Start
button to begin calibration. The rest depends on the
instrument setting:
Automatic Acceptance
.
When
Automatic Acceptance
is in use (checked), MC6-T accepts
calibration point automatically as follows:
1.
MC6-T uses the
Max. Point Deviation
value to see if the input signal
is close enough to the next calibration point.
2.
When close enough, MC6-T checks the signal stability to decide
whether the readings can be saved or not.
3.
When signal stability is reached, a timer counts down as set in
Point
Delay
and then the readings are saved only if the signal stability is
Figure 84: Calibration with Automatic
. If a signal becomes unstable, MC6-T returns to
phase 2.