118 - Communicator
Warnings
Warning:
Configuring/calibrating an instrument that is part of a
live segment is possible in HART and FOUNDATION Fieldbus.
When working in a live segment, first make sure, the control loop
the instrument is part of is set to manual. Follow the guidelines
available in the instrument's own manual.
Beamex cannot be held responsible for any damages caused by
connecting MC6-T to a live factory fieldbus segment.
Warning:
Using MC6-T to change the parameters of an
instrument may result in discrepancies: A fieldbus host control
system may mirror all instrument parameters in its permanent
database. In such a case, when returning an instrument
with changed parameters to a live segment, ensure that the
parameters are also available in the control system's permanent
database. Also verify that the new parameters do not result in an
unstable control loop.
Connections
When communicating with Digital Communication Instruments, the
connection differs depending on whether MC6-T's internal power supply is
used or not.
The
Figure 114: Using MC6-T's internal power supply
presents the
connections when
MC6-T's internal power supply is used
. MC6-T's
power supply can safely provide supply power to a single instrument. If
connecting to a HART/fieldbus segment containing several instruments,
consider using an external power supply as presented further on.