Running a performance check
Introduction
This topic describes how to run a performance check as part of
quality control.
Overview
The CS&T application is designed to monitor performance on a
daily basis and to optimize laser delay.
Running a performance check on a regular basis provides a
standard for monitoring changes in performance due to
degradation of laser power, aging of PMTs, and other potential
cytometer service issues. Performance results are also affected by
fluidics performance. We strongly recommend following the
fluidics maintenance procedures as described in
Considerations
Some BP filters might not be normalized to CS&T settings. In this
case, CS&T will generate Qr and Br numbers that are not
comparable from instrument to instrument. They are however, still
trackable on one cytometer. Part of the process for optimizing
cytometer settings includes verifying PMT voltages set by CS&T
for all parameters. Carefully examine any channel with a non-
CS&T normalized filter.
Before you begin
Define the performance baseline for any configuration before
running a performance check. See
.
Summary of Contents for FACSymphony A3
Page 4: ......
Page 8: ...User s Guide 8...
Page 14: ...This page intentionally left blank...
Page 46: ...This page intentionally left blank...
Page 86: ...This page intentionally left blank...
Page 97: ...Chapter 6 Recording and analyzing data 97 e Click OK 12 Print the analysis...
Page 98: ...User s Guide 98 Your global worksheet analysis objects should look like the following...