Fluidics system
Introduction
The fluidics system carries the sample out of the sample tube and into the interrogation region of the flow cell.
Cells are carried in the sample core stream in single file and measured individually.
Note:
If your system has the HTS option, see the
BD® High Throughput Sampler User's Guide
for additional
information on the sample flow rates and fluidic controls.
System indicators
There are two system indicators (System Status and Activity) on the control panel as well as Sheath Tank Air
Pressure, Sheath, and Waste indicators on the graphic display.
l
System Status bar.
Shows the status of the system by displaying a green, yellow, or red LED indicator. The
following table describes the LED indicators, conditions that trigger them, and any action that must be
taken.
LED
color
Status
Action
Condition
Green
Good
None
Good
Yellow
Sheath and/or waste tanks need
attention.
Check tank levels
Sheath low and/or waste
nearly full
Red
Take immediate action.
l
Empty waste tank
l
Fill sheath tank
l
Check air pressure
Sheath empty and/or
waste full
Air pressure is out of
specification
System status is also displayed on the Status screen. See
for a description of the
Status screen.
l
Activity.
Shows whether the cytometer power is on and the status of acquisition. The following table
describes the indicator LEDs, and status that triggers them.
Indicator
LED color
Status
Steady pulse
blue
Cytometer is powered on
Fluctuates
blue
Forward scatter is registering cells or particles
Fluid control
The three fluid buttons (RUN, STANDBY, and PRIME) control the fluidics subsystem.
l
RUN
Pressurizes the sample tube to transport the sample through the sample injection tube and into the
flow cell.
The RUN button is green when the sample tube is on and the support arm is centered and the sample tube is
pressurized. When the tube support arm is moved left or right to remove a sample tube, the cytometer
switches to an automatic standby status to conserve sheath fluid, and the RUN button changes to orange.
16
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...