Chapter 2: BD FACS Loader Option
43
Cleaning External Surfaces
To keep the Loader free from salt buildup, wipe down all external surfaces that
have been exposed to sheath fluid.
Materials
•
BD FACSClean solution
•
DI water
•
clean, lint-free cloths or disposable wipes
Procedure
1
Switch off the cytometer power and unplug the AC power cord from the
wall socket.
2
Wipe all accessible surfaces with BD FACSClean solution.
Ensure that the optical sensors are cleaned. Saline deposits can develop on
the sensors, causing tubes in the carousel to be missed.
3
Wet a fresh cloth with DI water and wipe all exposed surfaces to prevent
corrosion.
4
Wipe all exposed surfaces with a clean, dry cloth.
To avoid potential shock, always switch off the power and unplug the AC
power cord before you begin cleaning.