Appendix A: Technical Overview
115
Figure A-9
Light pathway around an octagon (4-2 configuration)
Behind the LP mirror, a 780/60 BP filter admits light from 750 to 810 nm and
substantially blocks other wavelengths. The light that finally reaches PMT A will
be from dyes such as PE-Cy7 that emit in this range.
750–810 nm
670-735 nm
515–545 nm
564–606 nm