BD FACSPresto Near-Patient CD4 Counter Instructions For Use
48
7. Press the
Upper Limit CD4 Percentage
field on the right and
enter a value.
8. Press the
Lot Number
field and enter the lot number for the
process control (1–10 characters).
9. Press
Accept
.
10. Press the
Expiration Date
field and press
Up
or
Down
to set
the year, month, and day.
11. Press
Accept
.
The Expiration Date screen closes.
12. Press
Accept
on the
CD4PC Low Settings
screen to save all the
entered data for the low count process control.
The CD4PC Low Settings screen closes.
13. Press
CD4PC N
on the
Process Control Settings
screen.
14. Repeat steps 4 to 12 for the normal count process control.
Defining Hb
process control
settings
To define the Hb process control settings:
1. Press
HbPC L1
on the
Process Control Settings
screen.