BDDF00160 Issue 1
23/40
Alaris™ GW Volumetric Pumps
Configurable Options Sheet
Configurable Options Sheet
w
The default settings are configurable as displayed in brackets in the table below. Each of the configurable options
has a code which must only be altered by Qualified Service Personnel with reference to the technical service manual
(TSM) for this product (Technical Service Manual reference: BDTM00006).
Description
Range
Default
Setting
Enable VTBI / time infusions
(On / OFF)
OFF
Maximum priming volume
(OFF, 1 to 40 ml)
40ml
Clear infusion parameters to zero on
power - up
(On / OFF)
OFF
Maximum VTBI in MICRO Mode
(0 .1 to 999 ml)
999ml
Bolus rate
(1 to 999ml/h)
400ml/h
Maximum bolus volume
(OFF, 1 to 99ml)
5ml
Keep vein open rate
(OFF, 1 .0 to 5 .0 ml/h)
5 .0ml/h
Air in line alarm volume - single bubble
(50,100, 250, 500µL)
100µl
Enable secondary infusion capability
(On / OFF)
OFF
Default occlusion pressure on power - up (Lo (250mmHg),Nor (350mmHg), HI (500mmHg)) HI
Alarm volume level
(1 - 7)
4
Enable Micro mode
(On / OFF)
OFF
Maximum infusion rate
(1 - 999 ml/h)
999ml/h
Enable ASCII mode for communications
(On / OFF)
OFF
Enable Odd parity for communications
(On / OFF)
OFF
Set Pump address for communications
(1 - 250)
1
Flow Sensor Connection Mode**
(AUTO/On)
On
Set - up of current time and date
(00:00 to 23:59)
(01/01/00 to 31/12/99)
N/A
Language selection
(EnGL, FrAn, dEut, ItAL, ESPA, SE, nEd)*
EnGL
IrDA Communications Selection
(On / OFF)
On
Nurse Call Activation High Enabled
(On / OFF)
On
Drops per ml of fluid
(1 to 200)
20
Enable Automatic Set Detection and Anti-
bolus Function***
(On / OFF)
OFF
Silent Mode
(On / OFF)
OFF
User select mode options
Pressure limit Enabled
Alarm volume Enabled
Timed infusions Enabled
Micro infusions Enabled
(On / OFF)
(On / OFF)
(On / OFF)
(On / OFF)
OFF
OFF
OFF
OFF
Flow sensor sensitivity level
(Nor, Hi)
Nor
*EnGL - English, FrAn - French, dEut - German, ItAL - Italian, ESPA - Spanish, SE - Swedish, nEd - Dutch.
** If infusion sets without anti-siphon valves are being used, it is recommended to change the flow sensor connection mode setting to
On
. With
On
selected, the Alaris GW 800 Volumetric Pump will not operate unless a flow sensor is connected.
*** When using dedicated BD Infusion sets (refer to Compatible Dedicated Infusion Sets section) ensure that the set detection mode is
On
(Refer
to Automatic Set Detection section for more details). Align all Alaris GW 800 Volumetric Pumps in your hospital to the same set detection mode
(ie. all
On
or all
OFF
) to avoid potential confusion amongst users as to the mode in which the Pump is operating under.
Serial Number
Software Version
Configured by
Date
Approved by
Date