Alaris™ GP (Guardrails™) Volumetric Pump
Prompts
BDDF00535
Issue 1
38/56
Prompts
Prompts are indicated by an audible alarm and message, they cannot be silenced and do not have a visual indicator .
Display
Infusion Status
Cause(s)
Action
SET VTBI
Pump on hold
No VTBI / flow sensor .
• Set VTBI or fit flow sensor .
LOCKED
Infusion continues
Rate change attempted
whilst locked .
• Unlock rate to adjust infusion
settings .
Advisories
Display
Infusion
Status
Cause(s)
Action
DOSE WOULD EXCEED
Pump on hold (If
titration is disabled)
Infusion continues
(If titrating)
Infusion rate set exceeds a
Guardrails soft limit .
• Check infusion setting .
• To confirm
CONFIRM DRUG?
press
YES
.
• To deny
CONFIRM DRUG?
press
NO
.
DOSE UNDER
Pump on hold (If
titration is disabled)
Infusion continues
(If titrating)
Infusion rate/dose rate set is
under a Guardrails soft limit .
• Check infusion setting .
• To confirm
CONFIRM DRUG?
press
YES
.
• To deny
CONFIRM DRUG?
press
NO
.
DOSE NOT PERMITTED
Pump on hold (If
titration is disabled)
Infusion continues
(If titrating)
Dose rate entered is greater
than the dose rate hard
maximum set .
• Check infusion setting and adjust
to appropriate required rate .
RATE NOT PERMITTED
Pump on hold (If
titration is disabled)
Infusion continues
(If titrating)
Infusion rate set exceeds a
hard limit .
• Check infusion setting and adjust
to appropriate required rate .
CONCENTRATION NOT
PERMITTED
Pump on hold
Concentration set exceeds
hard max limit, or is under
hard minimum limit .
• Check concentration and adjust
to a more appropriate amount .
WEIGHT ABOVE LIMIT
Pump on hold
Patient weight set exceeds a
Guardrails soft limit .
• Check weight setting .
• To confirm
CONFIRM?
press
YES
.
• To deny
CONFIRM?
press
NO
.
WEIGHT BELOW LIMIT
Pump on hold
Patient weight set is under a
Guardrails soft limit .
• Check weight setting .
• To confirm
CONFIRM?
press
YES
.
• To deny
CONFIRM?
press
NO
.
BOLUS DOSE NOT PERMITTED
Infusion continues
Bolus dose rate entered is
greater than the bolus dose
rate hard maximum limit .
• Check the bolus setting and
adjust to a more appropriate
dose .
BOLUS DOSE WOULD EXCEED
Infusion continues
Bolus dose rate set exceeds a
Guardrails soft limit .
• Check the bolus setting .
• To confirm
OVERRIDE LIMIT?
press
YES
.
• To deny
OVERRIDE LIMIT?
press
NO
.
BOLUS DOSE UNDER
Infusion continues
Bolus dose rate set is under a
Guardrails soft limit .
• Check the bolus setting .
• To confirm
OVERRIDE LIMIT?
press
YES
.
• To deny
OVERRIDE LIMIT?
press
NO
.