Chapter 2: Routine Maintenance
Advisor
®
Service Manual
2-9
Dielectric Withstand - Hi-pot Test
The following test described is NOT recommended to be performed as part of a comprehensive preventive
maintenance program.
CAUTIon! Dielectric Withstand Voltage / Hi-pot Testing is degradable to the monitor. Performing this test
may result in monitor failure.
Install shorting plugs in all patient connectors.
1.
Attach the Advisor
2.
®
unit to the Safety Analyzer by connecting the “red” lead to both prongs on the unit’s
power cord, and the “black” lead to the unit’s enclosure (shaft of the rotary encoder, remove knob for test).
Verify that no breakdown occurs at 1500 VAC for a period not more than 1 minute 6 seconds or less than 1
3.
minute.
Move the “black” lead to the shorting plug on the ECG connector.
4.
Verify that no breakdown occurs at 4000 VAC for a period not more than 3 seconds or less than 2 seconds.
5.
Repeat steps 4 and 5 for the oximeter, both IP channels (if installed), and both Temp channels (if installed).
6.
Move the “black” lead to the shorting plug on the serial port (if installed).
7.
Verify that no breakdown occurs at 1500 VAC for a period not more than 1 minute 6 seconds or less than 1
8.
minute.
Summary of Contents for Advisor
Page 2: ......
Page 6: ...Table of Contents iv Advisor Service Manual This page is intentionally left blank ...
Page 8: ...Revision History vi Advisor Service Manual This page is intentionally left blank ...
Page 43: ...Chapter 4 Diagrams Advisor Service Manual 4 1 Chapter 4 Diagrams ...
Page 44: ...Chapter 4 Diagrams 4 2 Advisor Service Manual ...
Page 45: ...Chapter 4 Diagrams Advisor Service Manual 4 3 ...
Page 46: ...Chapter 4 Diagrams 4 4 Advisor Service Manual ...
Page 47: ...Chapter 4 Diagrams Advisor Service Manual 4 5 ...
Page 48: ...Chapter 4 Diagrams 4 6 Advisor Service Manual ...
Page 49: ...Chapter 4 Diagrams Advisor Service Manual 4 7 ...
Page 50: ...Chapter 4 Diagrams 4 8 Advisor Service Manual ...
Page 56: ...Appendix App 2 Advisor Service Manual This page is intentionally left blank ...
Page 57: ......