![bbraun 8717070 Service Manual Download Page 23](http://html1.mh-extra.com/html/bbraun/8717070/8717070_service-manual_2714001023.webp)
1 Overview of the Device
Infusomat® compact
plus
P 1.0
1 - 7
EN
For internal use, only
FUNCTIONAL TEST
*
Danger of injury to the patient!
Device must not be connected to a patient.
Observe the Instructions for Use when performing the test.
* Do not use the device if the functional test fails. Contact your local representative for replacement or repair.
WARNING
!
Item
OK
Item
OK
1.
Completeness, Integrity
3.
Functional Test (see Instructions for Use)
1.1 Instructions for Use is available.
❒
3.1 Open pump door, insert a primed line, close pump
door.
❒
1.2 Final Inspection Sheet is available.
❒
1.3 No mechanical damage is found on Infusomat®
compact
plus
P.
❒
3.2 Open roller clamp, confirm line. "Rate" menu appears
after successful self-test.
❒
1.4 Type plate is present.
❒
3.3 Deny priming if this message appears (configurable).
❒
1.5 Read serial number from type plate and enter this
below on this page.
❒
3.4 Select a rate of 1.1 ml/h and start Infusomat®
compact
plus
P: Green LED lights up, arrows move from
right to left.
❒
1.6 Seal is present and undamaged.
❒
1.7 Infusomat® compact
plus
P is clean.
❒
3.5 Administer a bolus by preselecting a volume of 3 ml –
bolus is administered, volume counter counts up.
❒
2.
Switching on and Self-test
3.6 Close roller clamp – red LED lights up, alarm text is
displayed.
❒
2.1 Switch on Infusomat® compact
plus
P.
❒
2.2 Display shows "Self test".
❒
3.7 Confirm alarm by pressing the "OK" button and restart
Infusomat® compact
plus
P.
❒
2.3 Two signal tones are audible.
❒
2.4 Red LED flashes twice.
Green LED flashes once.
❒
3.8 Press "Menu" button – the "Main Menu" opens.
❒
3.9 Press "Info" button – therapy information is given.
❒
2.5 Is the battery charged?
If not: connect device to mains power supply.
❒
4.
Finalizing the Test
2.6 Is the required language displayed?
If not: select language.
❒
4.1 Remove line from Infusomat® compact
plus
P and
dispose properly. Switch device off.
❒
4.2 Place Instructions for Use with Infusomat®
compact
plus
P.
❒
4.3 Complete this form and file properly.
❒
4.4 Also file the Final Inspection Sheet.
❒
Functional Test performed at Infusomat® compact
plus
P with serial number:
Infusomat® compact
plus
P was handed over without any damage and is in proper working order.
Date
Signature B. Braun
Signature Owner
Edit
Summary of Contents for 8717070
Page 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Page 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Page 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Page 105: ......