Page 5 of 28 

OTH-SEM250-0356 Rev D

 

 

WARNING:  Do not attempt to clean or disinfect a contaminated single-use sensor. 

 

WARNING: Should the device contact contaminated surfaces (for example, if it falls on 

the floor) it should be cleaned and disinfected before obtaining another patient reading. 

 

WARNING: Do not attempt to disassemble or otherwise modify the Provizio SEM 

Scanner as this can result in unintended hazards.  The Provizio SEM Scanner System can 

only be serviced by the manufacturer (Bruin Biometrics, BBI) or a BBI Authorized 

Service Center.  Please contact your BBI Product Specialist or Customer Service 

regarding any device that is not functioning correctly. 

 

WARNING: To avoid the risk of electric shock, this equipment must only be connected 

to a supply main with protective earth. 

 

WARNING: To avoid the risk of explosion, the Provizio SEM Scanner should not be 

used in the presence of flammable anesthetic agents.  

 

WARNING: The Provizio SEM Scanner S will not operate correctly without first 

installing a single-use sensor head. 

 

WARNING:  Ensure the is no foreign material in the Charging Base slot before placing 

the scanner in the Charging Base. 

1.2.

 

Cautions 

 

CAUTION:  Examine the Provizio SEM Scanner for damage or sharp edges prior to 

each use.  If any damage or sharp edges are found, please contact Customer Service or 

your Product Specialist.  Do not attempt to use the device. 

 

CAUTION: Ensure the power cord is plugged into the supply main and the device is 

fully charged prior to use. 

 

CAUTION:  Do not immerse the Provizio SEM Scanner.  Immersion may damage the 

scanner and prevent it functioning normally. 

2.

 

General 

2.1.

 

Purpose 

The purpose of this document is to describe the steps necessary to configure the Bruin 

Biometrics LLC (BBI) System in a Healthcare Information System (HIS) as well as support the 

system in use. 

Summary of Contents for Provizio SEM

Page 1: ...6 Rev D PROVIZIOTM SEM SCANNER SCANNER S GATEWAY SYSTEM SUPPORT MANUAL Bruin Biometrics LLC 10877 Wilshire Blvd Suite 1600 Los Angeles CA 90024 USA Phone 310 268 9494 E mail customerservice bruinbiome...

Page 2: ...2 8 1 Exclusions 7 2 8 2 Disclaimer of Additional Warranties 7 2 9 Table of Symbols 8 3 SETUP INSTRUCTIONS INSTALL VMWARE PLAYER 9 4 CREATE VIRTUAL MACHINE 10 5 WEB SERVER VIRTUAL MACHINE SETUP 12 6...

Page 3: ...CONFIGURATION 19 10 1 Install the Webserver 19 10 2 Navigate to the Gateway 19 10 3 Configure the Gateway 19 10 4 Assign roles 20 11 HANDHELD SUPPORT 21 11 1 Electromagnetic Environment 21 11 2 Elect...

Page 4: ...r of a physician or other licensed health practitioner 1 1 Warnings WARNING The standard of care should be followed for reducing the risk of developing pressure ulcers Readings from the Provizio SEM S...

Page 5: ...the risk of explosion the Provizio SEM Scanner should not be used in the presence of flammable anesthetic agents WARNING The Provizio SEM Scanner S will not operate correctly without first installing...

Page 6: ...w adverse effects using the scanner 2 8 Warranty and Disclaimer This manual is intended to provide instructions on the proper use of the SEM It should be used in conjunction with the appropriate train...

Page 7: ...reof a that have been subject to misuse neglect or accident b that have been damaged by causes external to the Warranted Product c that have been used in violation of the BBI Instructions for Use d on...

Page 8: ...ed Caution or warning Dispose of this equipment according to local regulations for electrical and electronic waste disposal Manufactured By Drip Proof Equipment IPX1 The Provizio SEM Scanner enclosure...

Page 9: ...he terms in the License Agreement and then select Next to continue 4 Select the checkbox next to Enhanced Keyboard Driver and then select Next to continue 5 Uncheck both checkboxes on the User Experie...

Page 10: ...o continue 10 Set the Maximum disk size GB to 60 0 GB a Leave the default setting Split virtual disk into multiple files b Select Next to continue 11 Select the Customize Hardware button a Edit the Me...

Page 11: ...e the CTRL ALT DEL key on VMware Player menu bar to bypass the lock screen 22 Enter the password to continue 23 When the Networks window is displayed select the Yes option 24 Right click the windows s...

Page 12: ...ework 4 7 ii HTTP Activation found in NET Framework 4 7 WCF Services 1 In the new dialog box Add features that are required for HTTP Activation select the Add Features button iii IIS Hostable Web Core...

Page 13: ...under Allow in the Permissions for IIS_IUSRS section e Select Apply f Select OK ii Select OK to continue 9 Select the Windows button in the bottom left corner and type inetmgr a Open the Internet Info...

Page 14: ...of 28 OTH SEM250 0356 Rev D 6 Additional Setup Steps 1 Copy the contents of PN XXXX BBI Socket Installer to the Web VM 2 Launch the BruinSocket Setup install file a Select Next b Select Next c Select...

Page 15: ...mb drive into the host 3 Navigate to the file location of the saved BBI_SEM250_WEB VM a Ie C Users Host Documents Virtual Machines BBI_SEM250_WEB 4 Copy the BBI_SEM250_WEB folder 5 Paste the BBI_SEM25...

Page 16: ...ase engine server by right clicking the database engine and selecting Restart 7 Once the restart is complete expand the Security folder in the database engine right click the Logins folder and select...

Page 17: ...ase Engine as used in the previous steps 4 Select SQL Server Authentication for Authentication 5 Enter the BBI SQL username and password and select Connect 6 If authentication was successful and the B...

Page 18: ...Active Directory Users and Computers 3 Identify and navigate to the desired forest to contain the new BBI LDAP server roles e g cpx local 4 Right click the Users folder and select New and then Group...

Page 19: ...dow f Enter the following command g telnet IP of LDAP server port of LDAP Server e g telnet 192 168 1 11 389 h If the command prompt displays a blank command prompt screen with the title Telnet IP of...

Page 20: ...st LDAP Connection button and verify that a dialog box is displayed that says Connection was successfully established 6 Select Next to continue 7 Confirm the values and select Finish to proceed 8 Logo...

Page 21: ...ns Test Compliance Electromagnetic Environment Guidance RF emissions CISPR11 Group 1 The Provizio SEM Scanner 250 device uses RF energy only for its internal function Therefore its RF emissions are ve...

Page 22: ...ltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 30 cycles 5 UT 95 dip in UT for 5s 5 UT 95 dip in UT...

Page 23: ...where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is recommended separation distance in meters m Field strengths from fixed RF tr...

Page 24: ...in Table 5 Table 5 Separation Distance Separation distance according to frequency of transmitter meters Rated maximum output power of transmitter W 150 kHz to 80 MHz P d 2 1 80 MHz to 800 MHz P d 2 1...

Page 25: ...atability 1 0 2 SEM Value units Storage The Provizio SEM Scanner should only be stored at temperatures ranging from 4 F 20 C to 113 F 45 C at 5 to 90 relative humidity non condensing Operating Conditi...

Page 26: ...ministrator for the correct URL for your institutions installation I cannot save settings to the SQL Server Ensure you have sufficient SQL Server privileges to update settings SQL settings were not ve...

Page 27: ...ails Re enter the SQL settings and re verify Ensure you have sufficient SQL Server privileges to update settings Telnet connection to the LDAP server fails Re enter the SQL settings and re verify Ensu...

Page 28: ...ts Refer to https sem scanner com patents for more information 2019 Bruin Biometrics LLC or its affiliates All rights reserved Bruin Biometrics LLC 10877 Wilshire Blvd Suite 1600 Los Angeles CA 90024...