216
Rev. C
Message
Probable Cause and Corrective Action
Insufficient Sample
Volume. The system
cannot complete
analysis. Touch
Continue to begin
the sequence to
clear the system.
Replace the sample
port when prompted.
The system cannot report results for the patient or QC sample
because it does not have enough sample to complete
analysis. This can occur if the sample device does not contain
enough sample, or if an obstruction prevents the system from
aspirating enough sample for analysis. The Insufficient
Sample message appears in the events log.
1. Touch
Continue
to remove the message, then replace the
sample port when prompted. Refer to
Replacing the Sample
Port
in
Replacing Components
.
2. If prompted, replace the cartridges as described in
Replacing the Measurement and Wash/Waste Cartridges
in
Section 3
in the
Rapidpoint 400 Series Operator’s Manual
.
3. Ensure that the sample device you use contains sufficient
sample.
Refer to
Collecting Patient Samples
in Section 1
in the
Rapidpoint 400 Series Operator’s Manual
to identify the
minimum sample volume for the sample device that you are
using.
Service Solutions
Perform service solutions only after all other corrective actions
have been completed.
If cartridge replacement does not solve the problem, view the
events log for errors that may be causing the problem.
Continue with the associated service solutions.
M Cartridge Expired
The measurement cartridge has exceeded its shelf life or its
operating life or is depleted.
Refer to
Replacing the Measurement and Wash/Waste
Cartridges
in Section 3
in the
Rapidpoint 400 Series
Operator’s Manual
to replace both cartridges.
M Cartridge Not Valid The system detects that an expired measurement cartridge
was installed, that a used measurement cartridge was
reinstalled, or that a measurement cartridge was installed
without using the prompted method that you begin from the
Status screen.
The system is unable to use the cartridge.
Refer to
Replacing the Measurement and Wash/Waste
Cartridges
in Section 3 in the
Rapidpoint 400 Series
Operator’s Manual
to replace both cartridges.
(Continued)
Summary of Contents for RapidPoint 400 Series
Page 1: ...Service Manual 105953 Rev C 2003 02...
Page 29: ...Rev C Mechanical Descriptions 29 Rapidpoint 400 Series System left side 1 Diskette drive...
Page 62: ...62 Rev C...
Page 112: ...112 Rev C...
Page 114: ...114 Rev C...
Page 116: ...116 Rev C Rapidpoint 400 Series Systems Menu Map...
Page 117: ...Rev C System Diagrams and Cable Drawings 117 Fluidic Diagrams...
Page 118: ...118 Rev C...
Page 119: ...Rev C System Diagrams and Cable Drawings 119...
Page 120: ...120 Rev C...
Page 121: ...Rev C System Diagrams and Cable Drawings 121...
Page 122: ...122 Rev C...
Page 123: ...Rev C System Diagrams and Cable Drawings 123...
Page 124: ...124 Rev C...
Page 125: ...Rev C System Diagrams and Cable Drawings 125...
Page 126: ...126 Rev C...
Page 127: ...Rev C System Diagrams and Cable Drawings 127...
Page 128: ...128 Rev C Rapidpoint 400 Series Systems Electronic Block Text Diagram...
Page 131: ...Rev C System Diagrams and Cable Drawings 131 D C Power Harness Assembly 113836...
Page 132: ...132 Rev C Diskette Drive Cable Assembly 570255 Diskette Power Harness 570254...
Page 134: ...134 Rev C Inboard 2 Harness 113523 LCD Backlight Cable Assembly 108771...
Page 135: ...Rev C System Diagrams and Cable Drawings 135 Preamp Module Cable Assembly 570269...
Page 136: ...136 Rev C UIP Display Cable Assembly 570263 UIP LPT1 COM2 Cable Assembly 570267...
Page 138: ...138 Rev C Flex Circuit Display 570516 100 Pin Cable Assembly 111326 Patch Cord Cable 122254...
Page 139: ...Rev C System Diagrams and Cable Drawings 139 Fiber Bundle Assembly 1315271...
Page 140: ...140 Rev C...
Page 142: ...142 Rev C...
Page 148: ...148 Rev C...
Page 154: ...154 Rev C...
Page 258: ...258 Rev C...
Page 260: ...260 Rev C...
Page 262: ...262 Rev C Handle Pawl...
Page 263: ...Rev C Illustrated Parts Lists 263 Wash Waste Switch Rubber Feet...
Page 264: ...264 Rev C Display Display Bracket...
Page 267: ...Rev C Illustrated Parts Lists 267 Front Door Door Hook...
Page 268: ...268 Rev C Cartridge Interface Assembly Front Door Switch...
Page 271: ...Rev C Illustrated Parts Lists 271 Diskette Drive Diskette Drive Cable Shield...
Page 273: ...Rev C Illustrated Parts Lists 273 Connector UIP PCB UIP PCB with Cables...
Page 274: ...274 Rev C Main PCB Main PCB Attached to Frame...
Page 276: ...276 Rev C AQC Latch Assembly Front View AQC Latch Assembly Rear View 1 AQC latch assembly...
Page 277: ...Rev C Illustrated Parts Lists 277 Cover...
Page 282: ...282 Rev C Pump Housing Peri pump Motor and Mount...
Page 283: ...Rev C Illustrated Parts Lists 283 Peri pumps Spring Plate...
Page 284: ...284 Rev C Connector Block Cartridge Handle...
Page 285: ...Rev C Illustrated Parts Lists 285 Preamp Cable Cover...
Page 289: ...Rev C Illustrated Parts Lists 289 Power Module Frame Power Entry with Switch...
Page 290: ...290 Rev C Rear Interconnect Board Fan...
Page 291: ...Rev C Illustrated Parts Lists 291 Power Supply Patch Cord with Ferrite Filter...
Page 292: ...292 Rev C Hard Drive...
Page 296: ...296 Rev C Neon Board Assembly Part Number 131526...
Page 298: ...298 Rev C...
Page 302: ...302 Rev C...
Page 351: ...Rev C Replacing Components 351 a Remove the four thread forming screws from the motor mount...
Page 380: ...380 Rev C...
Page 381: ...Rev C Spare Parts 381 Spare Parts Spare Parts Listing 383...
Page 382: ...382 Rev C...
Page 386: ...386 Rev C...
Page 400: ...400 Rev C...
Page 426: ...426 Rev C...
Page 432: ...432 Rev C...
Page 438: ...438 Rev C...
Page 458: ...458 Rev C...