7-2-6
Barcode Test
7-2-6-1
From Service Test, press
ENT
to access the six tests available. The third displayed test, Barcode
Test, may be accessed by pressing
ENT
when this title appears (see Figure 7-3). If
ESC
is
pressed, the Service Test menu appears. If
MENU
is pressed, the unit goes on to the next test.
7-2-6-2
Press
ENT
and display indicates
揃
arcode Test.?Once in this menu two choices are available:
1.
press
ENT
to take a barcode reading
2.
press
ADJ
to move the crossfeed arm
7-2-6-3
Press
ENT
from the Barcode Test display and the display changes to read
揃
arcode Reading.?
This screen takes multiple readings of bar code labels on tubes while they are in front of the reader.
7-2-6-4
If a bar code label is detected by the reader, the bar code number will flash repeatedly on the
display. As long as a bar code label can be read, the number read will continue flashing. If the
display remains stationary and unchanged, even with a bar code number on it, no label was
detected. If the letters
NG
?(No Good) appear after the bar code number, the instrument had
difficulty in reading the label. Press
STOP
to end the readings and return to the Barcode Test
display.
NOTE:
The barcode reader also has a green
揙
K?LED and red
揘
G?LED, used to test the readability of
the barcode reader and labels.
7-2-6-5
Press
ADJ
from the Barcode Test display and the display changes to read
揚
osition Adjust.? This
screen permits adjusting or moving the position of a tube rack to the left
(LEFT)
or right
(RIGHT)
by
moving the crossfeed arm (see item 32 on page 10.28) left or right.
7-2-6-6
The amount of movement may be set from 0.1 to 5.0mm. If it is set to 1.0mm, each time softkey
LEFT
is pressed, the crossfeed arm is moved 1mm to the left and moves a rack with it. Each time
softkey
RIGHT
is pressed, the crossfeed arm is moved 1mm to the right. The rack must then be
pushed by hand to the right to touch the crossfeed arm in its new position, since the arm cannot
損
ull?the rack backwards.
7-2-6-7
These tests are used to test the performance of the PH4 Tube Sensor, barcode reader and barcode
labels:
To use these tests to examine the PH4 Tube Sensor, see paragraph 8-6 on page 8.7.
To test a barcode reader and barcode labels, see paragraph 8-7 on pages 8.8 and 8.9.
Revised 6/98 page 7.7
7-6
Summary of Contents for Clinitek Atlas
Page 1: ...Clinitek Atlas Rack Sample Handler Service Manual Cover...
Page 14: ...CHAPTER FOUR OPERATIONS PROCEDURES SUMMARY Figure 4 1 Main Menu Operation Block Diagram 4 1...
Page 20: ...Figure 6 2 Cross Feed Assembly 6 3...
Page 22: ...Figure 6 4 Power Module Figure 6 5 The Interconnect PC Board and R1 Relay 6 5...
Page 24: ...Figure 6 6 The Card Cage Assembly 6 7...
Page 26: ...Figure 6 8 The CPU PC Board Figure 6 9 The I O PC Board 6 9...
Page 37: ...Figure 7 1 Service Mode Block Diagram 7 3...
Page 39: ...Figure 7 2 Without Tube Run Routine and With Tube Run Routine 7 5...
Page 41: ...Figure 7 3 Barcode Test 7 7...
Page 46: ...Figure 7 5 Output Test Figure 7 6 Display Test 7 11...
Page 66: ...Figure 8 5 Adjusting Tube Sensor PH4 8 6...
Page 72: ...CHAPTER NINE REPAIR REPLACEMENT 9 1...
Page 73: ...Figure 9 1 Removing the Covers Part One Figure 9 2 Removing the Covers Part Two 9 2...
Page 75: ...Figure 9 3 The Return Feed Assembly 9 4...
Page 77: ...Figure 9 4 The Rack Feed Assemblies In Feed Out Feed 9 6...
Page 79: ...Figure 9 4 The Rack Feed Assemblies In Feed Out Feed 9 8...
Page 81: ...Figure 9 4 The Rack Feed Assemblies In Feed Out Feed 9 10...
Page 83: ...Figure 9 5 The Main Chassis Assembly 9 12...
Page 85: ...Setting the Timing Belt Deflection Figure 9 7 The Cross Feed Assembly 9 14...
Page 87: ...Figure 9 8 Proper Installation of Wire Cables on Winding Drums 9 16...
Page 89: ...Figure 9 10 Removing and Replacing the Transfer Plate 9 18...
Page 91: ...CHAPTER TEN PARTS BREAKDOWN ILLUSTRATED ASSEMBLIES 10 1...
Page 94: ...Figure 10 1 Stat Tube Holder Assembly 10 4...
Page 96: ...Figure 10 2 Stat Tube Holder Covers 10 6...
Page 98: ...Figure 10 3 Return Feed Assembly 10 8...
Page 101: ...Figure 10 4 Rack Feed In Feed Out Feed Assembly 10 10...
Page 103: ...10 4 32 Cap Screw M4 x 18 4 10 4 33 Spring Washer 4 1 10 4 34 Plain Washer 4 4 10 11 a...
Page 104: ...Figure 10 4 Rack Feed In Feed Out Feed Assembly 10 12...
Page 106: ...Figure 10 5 Display Assembly 10 14...
Page 108: ...Figure 10 6 Power Module Switch Box 10 16...
Page 118: ...Figure 10 15 Main Chassis Assembly 10 26...
Page 120: ...Figure 10 16 Cross Feed Assembly 10 28...
Page 124: ...Figure 10 19 Main Chassis Unit Frames 10 32...
Page 126: ...Figure 10 20 Frame Assembly 10 34...
Page 132: ...Figure 10 25 Additional Parts Part 1 10 40...
Page 134: ...Figure 10 26 Additional Parts Part 2 10 42...
Page 136: ...Figure 10 27 Additional Parts Part 3 10 44...
Page 138: ...Figure 10 28 Additional Parts Part 4 10 46...
Page 140: ...Figure 10 29 Additional Parts Part 5 10 48...
Page 142: ...Figure 10 30 Covers Part 1 10 50...
Page 144: ...Figure 10 31 Covers Part 2 10 52...
Page 154: ...11 6 a...
Page 157: ...Figure 2 Installing Drill Fixture and Drilling Holes 11 9...
Page 163: ...11 14 a...
Page 168: ...Figure 3 Setting the Belt Tension 11 19...
Page 170: ...Figure 4 Setting the JP1 Jumper on the Power Driver PC Board 11 21...
Page 172: ...SCHEMATICS 12 2...
Page 173: ...Figure 1 System Block Diagram 12 3...
Page 174: ...Figure 2 AC DC Power Circuit 12 5...
Page 175: ...Figure 3 Cables and Assembly Wiring Number 1 12 7...
Page 176: ...Figure 4 Cables and Assembly Wiring Number 2 12 9...
Page 177: ...Figure 5 Cables and Assembly Wiring Number 3 12 11...
Page 178: ...Figure 6 Cables and Assembly Wiring Number 4 12 13...
Page 179: ...Figure 7 Atlas Interconnect PC Board 12 15...
Page 180: ...Figure 8 Cable Assemblies Number 1 12 17...
Page 181: ...Figure 9 Cable Assemblies Number 2 12 19...
Page 182: ...Figure 10 Cable Assemblies Number 3 12 21...
Page 183: ...Figure 11 Driver PC Board 12 23...