FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
8-27
Removal/Replacement Procedures
Disassembly/Reassembly
5.
Install the display board. Tightening torque for the five screws
(Figure 11-2 , item 15) securing the board is 1.5 kgf-cm (1.3 in-lb).
6.
Reassemble in reverse order. Be sure to remove any dirt or dust on the
LCD windows before tightening the sensor board to the front housing.
Be sure to reconnect all the removed connectors.
7.
Perform the Operational Checkout procedures in Chapter 10,
“Operational Checkout”.
Replacement of Audible Alarm/Alert Board
1.
Separate the front and rear housings in accordance with “Separation of
Front and Rear Housings” on page 8-5.
2.
Remove one screw (Figure 11-3 , item 21) securing the audible alarm/
alert board (item 3) to the rear housing.
3.
Remove the board and unplug connector CN753.
4.
Reassemble in reverse order. Tightening torque for item 21 is 7 kgf-cm
(6.1 in-lb).
5.
Perform the Operational Checkout procedures in Chapter 10,
“Operational Checkout”.
Replacement of Backup Audible Alarm/Alert Board
1.
Separate the front and rear housings in accordance with “Separation of
Front and Rear Housings” on page 8-5.
2.
Unplug connectors CN751 and CN752.
3.
Remove one screw (Figure 11-3 , item 21) securing the backup audible
alarm/alert board (item 2) to the rear housing.
4.
Remove the board.
5.
Reassemble in reverse order. Tightening torque for item 21 is 7 kgf-cm
(6.1 in-lb).
6.
Perform the Operational Checkout procedures in Chapter 10,
“Operational Checkout”.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...