Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 62
53007410 Rev. C
Priming the Inlets and Verifying the Setup
Usually, a cosigner must log in and verify the setup. Refer to your facility’s protocol.
WARNING
It is important to have a cosigner independently verify the setup, to help ensure that the
first user attached each ingredient’s inlet to the correct port. Incorrect setup could result
in patient harm.
IMPORTANT!
The cosignature option requires Verification permissions. For more information
about user groups and permissions, refer to Setting Up the Users on Page 123. To require that a
cosigner verify the setup, refer to Cosignature on Page 120.
1.
Attach a calibration bag. Refer to Attaching the Calibration Bag on Page 73.
2.
If the
Cosignature required
message appears, the cosigner should:
a.
Tap
OK
.
Message
b.
Enter a
Login
name.
c.
Enter a
Password
.
d.
Tap
Log In
, then continue with the next steps.
Cosignature window