
Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 199
Issue / On-screen Text
Explanation
Suggested Actions
Text:
Formula Conflict:
Additional <
flush volume
>
mL of <
Universal
Ingredient
> required for
flush.
The formula does not
contain the minimum
required Universal
Ingredient.
WARNING
If you choose to
Increase Ingredient
Volume
, the clinical impact should be
considered. This option increases the
volume in the bag and is not part of the
original order.
Increasing the Universal Ingredient
volume may change the overall formula
ordered. Consult a pharmacist before
compounding.
IMPORTANT!
This function requires Formula Edit
permissions, which Baxter does not recommend
using. For more information about user groups and
permissions, refer to Setting Up the Users on Page
Tap one of these options:
•
Tap
Change UI To…
to change the Universal
Ingredient to one that has the minimum volume
required and does not have flush ingredient,
without increasing the ordered volume. Then
tap
OK
and continue with the steps below.
WARNING
A calibration bag must be used during
all Universal Ingredient flushes. You
must replace the original patient bag
with a calibration bag for the flush, then
reattach the original patient bag. If this
is not done, the patient bag could
contain an unintended volume and/or
ingredient.
1.
If a patient bag is attached, remove it. Refer to
Removing the Patient Bag on Page 90.
2.
Change the Universal Ingredient. Refer to
Changing the Universal Ingredient on Page 107,
starting at Step 2.
3.
Attach the patient bag. Refer to Attaching the
Patient Bag on Page 85.
4.
Repeat the compounding process.
Summary of Contents for ExactaMix 2400 Compounder
Page 2: ......