Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 117
Acceptable Weight Variances (%)
For
Final Solution
, enter the maximum acceptable difference between the expected and actual
weight of the compounded solution. If any compounded solution has a weight outside this
range, an alarm beeps and a message displays the results in red. The results also appear on the
MixCheck Report.
For
Individual Ingredient
, enter the maximum acceptable difference between the expected and
actual weight of each delivered ingredient. If any delivered ingredient has a weight outside this
range, a message appears. The compounder weighs only ingredient deliveries of 100 mL or
more.
Tip!
Baxter recommends entering 5% for the
Final Solution
and 5% for the
Individual
Ingredient
.
Manual Add
For
Max manual add volume
, enter the maximum volume allowed for a manual addition. If the
volume of the formula ingredient exceeds this amount, a message appears, with options to add
the ingredient manually or cancel compounding.
Flush Between UI Changes
For
Volume (mL)
, enter the final flush volume used to clear the common fluid pathway after
changing the Universal Ingredient.
NOTE:
A UI flush contains three deliveries with standard volumes of 50, 50 and 30 mL. Changing
the
Volume
setting changes only the last of the three deliveries.
Report Printer
Select the
printer used for printing reports.
MixCheck Data Export Printer
Select the printer used when sending MixCheck data to the DoseEdge system.
If
MixCheck Data Export
option (System tab) is enabled, application will display
MixCheck Data
Export Printer
in System Cont. tab.