4
Battery Charge Icon Descriptions
4-8
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Operating on Battery Power
The pump can be battery-powered in emergency situations and while
transporting patients. AC power should be used when not transporting
patients.
The battery charge icon is displayed at all times in the upper left part of
the screen, even when the device is operating on AC power (Figure
4-11). When the pump is unplugged and operating on battery power, the
battery charge icon alternates with the Plug In icon, and the approximate
battery operating time is displayed below the icon.
Charge the battery for at least 12 uninterrupted hours or until the battery
operating time displayed under the battery charge icon in the upper left
part of the screen is 3h 15m (for triple channel pumps) or 4h 00m (for
single channel pumps). A complete charge may take longer than
12 hours.
Battery time remaining adjusts based on flow rate; the batteries will
discharge at a faster rate at higher infusion rates. If the flow rate is
changed during an infusion, the battery time remaining will change
accordingly.
Battery Charge Icon Descriptions
See Table 4-1 for a list of battery charge icons and their meaning.
Battery time remaining is approximate.
Figure 4-11 Battery Charge Icon on
Main Display
Table 4-1
Battery Charge Icons and Descriptions
Icon
Description
The battery charge icon is displayed at all times in the upper left part of the
screen. The number of filled areas in the battery charge icon is an
approximate indication of the battery charge level. When the battery time
remaining is 80 to 100%, the icon contains three green bars as shown at left.
As the battery charge level decreases, the battery charge icon changes. When
the battery time remaining is 60 to 80%, the icon contains two green bars as
shown at left.
As the battery charge level continues to decrease, the battery charge icon
continues to change. When the battery time remaining is at 20 to 60%, the
icon contains two yellow bars as shown at left.
When the battery time remaining is less than 20%, the battery charge icon
contains one red bar as shown at left.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...