071960491
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
11-19
Quick Reference Guide
Unloading the Administration Set
Reset the Manual Tube Release as follows:
1.
Close the regulating clamp on the administration set.
Ensure
there is no administration set or foreign object in the tubing channel.
2.
Turn the release tab (Figure 11-16A) clockwise until it stops and
push the tab into its socket. For triple channel pumps: Repeat the
steps above as needed for additional channels.
3.
Press the
Done
soft key to clear the alarm.
Note:
If the MTR is used following a channel failure to remove the
administration set, the pump cannot be powered off until the MTR
has been reset. A Reset Manual Tube Release pop-up message
will be displayed.
Note:
If three unsuccessful attempts to reset the MTR are made, a
channel failure occurs. The pump cannot be used until the MTR
is reset and the pump is powered off and back on.