Chapter 9
071960491
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
9-1
Chapter 9
Technical Specifications
Pump Specifications
Component
Description
Device Type
Baxter product code 2M91617
Single Channel Shuttle Volumetric Infusion Pump
Baxter product code 2M91637
Triple Channel Shuttle Volumetric Infusion Pump
Administration Sets
Standard Baxter administration sets equipped with keyed slide clamps.
See “Recommended Administration Sets,” 4-15.
AC Power Requirements
100/120 VAC 50/60 Hz or 220-240 VAC 50/60 Hz.
Leakage Current
Less than 300 microamperes earth leakage (tested per UL 60601-1).
External Fuses
1.6 amp Type T (time delay) 250V.
Power Cord
Approximately 2.7 m (9 feet) long with integrally molded plug.
Battery Supply System
For single channel pumps: Fully charged new batteries typically provide
approximately 4 hours of operation at 100 mL/hr. Battery run time is dependent on
the channel flow rate and device/clinician interaction.
For triple channel pumps: Fully charged new batteries typically provide
approximately 2 hours of operation with all three channels running at 100 mL/hr.
Battery run time is dependent on the channel flow rate and device/clinician
interaction.
Internal charge system recharges batteries whenever pump is connected to an AC
outlet. See “Battery Care,” 7-3 for more information about the batteries.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...