Troubleshooting Alerts
8-22
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
X - Unconfirmed Primary
Program
X - Unconfirmed Secondary
Program
(no pump module alert message
)
With a primary infusion running, the
START
key was pressed prior to
confirming programming information.
Press the
Confirm Primary
or
Confirm Secondary
soft key
followed by the
START
key.
Table 8-2
Troubleshooting Alert Messages — continued
Alert Message
Cause
Recommended Action