Programming the Pump
51
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
The SIGMA Spectrum Infusion System is not intended to replace
clinician patient observation.
The Pump was not designed nor is it intended to detect infiltrations or
extravasation
WARNING
Confirm Drug Library.
Master Drug Library Administrators (MDLAs) should verify the
correct Drug Library is installed when deploying the Drug Library to
Pumps.
Master Drug Library Administrators (MDLAs) should verify the
Drug Library transfer is successful after deployment.
Users should verify the correct Drug Library is installed on the
deployed Pumps.
Before implementation, clinical users at each facility must
thoroughly test and validate their Drug Library per their facility’s
procedure to ensure configuration and workflow reflect clinical
practice.
CAUTION
Confirm Audio Operation.
Listen for beeps when pressing keys. If sound is not heard,
discontinue use of the Pump and refer servicing to qualified service
personnel at your facility or return the Pump to Baxter for service.
CAUTION
Confirm Display Operation.
Regularly observe the Pump’s display. If display abnormalities are
observed, discontinue use of the Pump and refer servicing to
qualified service personnel at your facility or return the Pump to
Baxter for service.
To program the Pump using DERS:
1. Press
ON/OFF
to power on the Pump.
2. The New Patient screen may display.
The Pump retains previously programmed infusion values up to 24 hours after power off. The
New Patient screen will display at power on to allow use of the previously programmed values
or allow a new infusion to be programmed.
3. At the New Patient screen:
Press the
yes
soft key to clear the previously programmed infusion values and allow
programming a new infusion.
NOTE:
If it is the same patient, but a new drug is required, press the
yes
soft key. As a
result, the Total given will be reset to zero.