Ipump Pain Management System Operator’s Manual
5-12
CD07-19-E4-766
Use the Initial Prescription
Action
Message Displayed
1.
At the
USE INITIAL Rx?
prompt, press the
⇑
key to display the options YES or
NO, then press
Enter
to select. (If necessary, refer to INITIAL Rx in the Glossary.)
If YES is selected, the pump will display the programmed values from the previous
initial prescription as default values. Carefully review each screen and press
Enter
to
accept, or modify the prescription if desired by entering different values (for example
for the mode, unit, and/or bolus), and then press
Enter
to accept the new value.
If NO is selected, the pump will display
SELECT MODE
, allowing you to program a
new prescription. Refer to page 5-13.
NOTE:
Rx is a symbol that represents a prescription for drugs or medical devices.
USE INITIAL Rx?
↑
NO
! WARNING !
To help prevent medication errors, Baxter recommends that both the
clinician programming the pump and another clinician check the accuracy
of prescription and programming information before the infusion is started.
2.
If necessary, press
History
to review the previous therapy history as described in the
“Reviewing the Therapy History,” 5-37.
3.
The pump returns to
USE INITIAL Rx?
following the history review.