Ipump Pain Management System Operator’s Manual
5-6
CD07-19-E4-766
Turn Off the Pump
The prescription and units cannot be changed unless the pump is turned off and then reprogrammed. If the
pump is turned off, the current prescription data will be lost.
Note:
If it is necessary to retain a patient’s history information, you can review and record the history data on the patient’s chart
or print a hard copy before proceeding with this procedure. See “Reviewing the Therapy History,” 5-37.
When turned back on, the pump reverts back to the initial prescription programmed for that patient. See
“Restarting the Infusion,” 5-33.
Action
Message Displayed
1.
If an infusion is in progress, press
STOP
twice within 1 second. This will stop
the infusion but will not turn off the pump.
KEY+CODE or KEY ONLY:
PUMP READY
PRESS START
CODE ONLY:
PUMP READY
START OR CLEAR
2.
Press the
ON/OFF
key twice within 1 second to turn the pump off.
Oct02_0719E4766.book Page 6 Friday, August 3, 2007 1:46 PM