Ipump Pain Management System Operator’s Manual
3-10
CD07-19-E4-766
Unlocking and Opening the Cover
1.
If the pump is attached to a pole, remove it as described in “Remove the Pump from the Pole,” 3-23.
2.
Place the pump face down.
3.
If the cover is locked, insert and push in the key, then rotate it one-quarter turn counterclockwise. Then,
open the cover. The key will remain in the lock whenever the cover is unlocked. To lock the cover, insert
and push in the key, then rotate it one-quarter turn clockwise (see Figure 3-5).
Removing or Changing the Cover
The highest level of drug security is attained by programming the combination of KEY+CODE and using the
cover. However, if the pump is configured for CODE ONLY
,
the cover may be removed completely.
1.
To remove or change the cover, unlock the cover (if necessary) and open it as described above.
2.
Place the pump face down.
3.
Detach the cover (and hinge cover, if removing the 250E bag cover) from the pump by removing the three
screws on the hinge assembly (see Figure 3-6). Store the cover and hinge cover for future use.
Note:
The hinge cover is for use with 250E bag covers only.
4.
Attach the new cover, if desired, by aligning the hinge assembly and replacing the three screws.