Always clean/disinfect the pump and PCA button after each use as follows:
Type of use
Action
If the device has been in an isolation area Select those agents from the list that both clean and disinfect.
If the device has been used on a patient
Clean/disinfect with an agent from the recommended list of cleaners before use on another
patient.
If spills occur or the device is dirty
Clean as quickly as possible to minimize any potential difficulties with the solutions
pooling and drying on the mechanism. If fluid enters the tubing channel, contact your local
Baxter Service Center (see “Authorized Service Centers,” 7-1).
! WARNING !
Liquids must not be allowed to enter the inside of the pump through the battery compartment,
or through the power, printer, or PCA ports. Liquid ingress into the pump may cause the pump
to fail or operate in an unintended manner during the current or future patient use.
CAUTION
The Ipump Pain Management System is not waterproof and should not be immersed in water.
Avoid getting liquids inside the pump. Air sensor functioning could be compromised or per-
manent damage may result. Do not use alcohol for cleaning.
CAUTION
Do not clean, disinfect, or sterilize any part of the device by autoclaving or with ethylene oxide
gas. Doing so may damage the device and void the warranty. Only external parts of the device
should be disinfected.