Ipump Pain Management System Operator’s Manual
6-15
CD07-19-E4-766
SET NOT PRIMED
PRESS ENTER
Note:
If no air is visibly
present in the
shorter end of the
tubing set, reload
the tubing set,
ensuring the shorter
end is not stretched.
Situation: Air is detected in the tubing during the upstream occlusion test.
Action:
1.
Press
Enter
. The pump will display the
START TO PRIME, ENTER TO
PROCEED
prompt.
! WARNING !
The tubing set MUST NOT be connected to the patient while priming.
2.
Press
START
:
If the pump is configured as KEY+CODE or CODE ONLY, enter the security
code at the
000 ENTER CODE
prompt, and press
Enter
.
If the pump is configured as KEY ONLY, unlock and lock the fluid bag cover at
the
UNLOCK THE COVER
and
LOCK THE COVER
prompts. Priming begins (see
“Prime the Tubing Set,” 5-16).
3.
After the tubing set is primed, press
Enter
for the
PUMP READY, START OR
CLEAR
prompt.
4.
Press
START
to run the upstream occlusion test again and begin infusing.
Table 6-2
Alarm Messages and Responses — continued
Alarm Message
Situation/Action
Oct02_0719E4766.book Page 15 Friday, August 3, 2007 1:46 PM