Ipump Pain Management System Operator’s Manual
6-11
CD07-19-E4-766
BATTERY IS
DEPLETED
Situation: The pump is running on battery power and the battery power remaining is
too low to continue.
Action:
1.
Replace the dead battery as soon as possible as described in “Installing and
Changing the Battery,” 3-2.
2.
If necessary, connect the pump to the AC adapter (see “Connecting the AC
Adapter,” 3-5).
CHECK TUBING
PLACEMENT
Situation: The tubing set is either improperly loaded or damaged.
Action:
1.
Ensure that the set is not damaged and is properly loaded. For information on
loading the tubing set, refer to “Preparing, Loading, and Changing the Tubing
Set and Fluid Bag,” 3-13.
NOTE:
If the tubing door is opened and closed, the upstream occlusion test will run when
START
is pressed. See “Upstream Occlusion Testing,” 5-18.
2.
If the alarm condition persists, a sensor failure may exist. Do not use the pump.
Take the pump out of service and contact your nearest authorized service center.
Table 6-2
Alarm Messages and Responses — continued
Alarm Message
Situation/Action