Ipump Pain Management System Operator’s Manual
6-5
CD07-19-E4-766
BATTERY IS
MISSING
Or
BATTERY MISSING
Situation: The pump is powered by the AC adapter and no battery is inserted.
If another action is occurring at the same time, the pump displays the message
BATTERY MISSING
on the second line of the screen.
Action: Follow the directions for replacing the battery under “Installing and Changing
the Battery,” 3-2.
BATTERY VOLTAGE
IS LOW
Situation: The pump has just completed the power on self-test and the battery power
is low. Three short tones are sounded.
NOTE:
If the battery voltage reaches a low level during an infusion, the
BATTERY VOLTAGE IS LOW
condition could become an ALARM.
Action:
1.
Press
Clear/Silence
during the alert to silence the audio for 2 minutes,
regardless of the Alert Silencing Time configuration setting.
2.
Replace the battery as soon as possible as specified in “Installing and Changing
the Battery,” 3-2.
Table 6-1
Alert Messages and Responses — continued
Alert Message
Situation/Action