Connecting the system
822-188-3
49 / 74
6.12.3 Connection with an HV connection cable with MC plug
1. Screw the HF filter onto the
HV IN
connection of the PD-TaD 80.
2. Screw the threaded pin (included in the standard delivery) to the HF filter.
3. Insert the MC plug-and-socket connection of the HV connection cable onto the threaded
pin and screw it on tight.
6.13 Connecting the system to the supply voltage
PD-TaD 80 integrated in the cable test van
Connect the cable test van to the supply voltage. Follow the instructions in the user
manual for the cable test van.
The
PWR
and
PD
LEDs on the PD-TaD 80 light up.
PD-TaD 80 as a standalone system
NOTICE
Too high or too low mains voltage
A low mains voltage adversely affects the function of the system, a high mains
voltage can cause damage.
Ensure that the supply voltage matches the specifications on the rating plate.
1. Measure the mains voltage with a voltmeter.
2. Compare the mains voltage with the specifications on the rating plate.
3. Make sure that the mains supply earth is not isolated from the station earth.
CAUTION
High electric voltage through potential increase.
Risk of personal injury due to
electric shock. Damage to property due to potential differences from mains input to the
housing.
4. Check whether the PD-TaD 80, VLF HV generator and, if the case may be, the laptop
are connected to the Power Box to receive a power supply.
5. Connect the Power Box to the mains voltage.
The green LED on the Power Box and the LEDs
PWR
and
PD
on the PD-TaD 80 light
up.