PD-TaD 62
Parallel partial discharge and dissipation factor measurement
822-187-3
61 / 114
During a parallel PD and dissipation factor measurement, the dissipation factor
measurement steps on the VLF HV generator and the PD measurement steps are
configured and controlled via the BAUR system software.
The dissipation factor measurement is carried out on the basis of templates. A template
offers the following information:
Program: Parameters for carrying out the diagnostic testing
Evaluation criteria
You have the option to configure templates for different diagnostics tasks and save them for
further use.
When creating templates you can combine existing programs and evaluations in any way or
create new programs and evaluations. A template in accordance with IEEE 400.2 is also
available.
These instructions describe how to set up a new template based on an existing program and
an existing evaluation, and on a template in accordance with IEEE 400.2. Further information
on the templates and how to create programs and evaluations can be found in the user
manual for the VLF HV generator in question.
9.1
Creating templates for dissipation factor measurements on
the VLF HV generator
9.1.1
Create new template for dissipation factor measurement
Where you can create a template
You can create templates for dissipation factor measurements on the VLF HV generator (see
description below).
How you can create a new template
You can create a new template in different ways:
completely new from scratch
Further information: Chapter
Creating a template newly from scratch
(on page 62)
on the basis of a template that you have created at an earlier point in time
For the new template a new program and a new evaluation can be created or already
existing programs and evaluations can be selected and, if necessary, adapted.
9
P
ARALLEL PARTIAL DISC HARGE AND DISSIPATIO N
FACTOR MEASUREMENT
Summary of Contents for PD-TaD 62
Page 26: ...Technical data PD TaD 62 26 114 822 187 3 4 TECHNICAL DATA ...