Maintenance
DTA 100 C
72 / 90
822-129-8
16.1 Cleaning the oil breakdown voltage tester
DANGER
Electrical voltage on device
Danger to life or risk of injury from electrical voltage.
Switch off the device before cleaning.
To disconnect the device completely from the mains voltage, pull
out the mains plug.
WARNING
Fire hazard posed by petroleum ether or other cleaning solvent
We recommend using cleaning solvent to clean the test chamber of the
oil breakdown voltage tester and the test vessel. The cleaning solvent
is highly flammable and in some circumstances may cause a fire.
Do not smoke when working with cleaning solvent.
Avoid naked flames.
NOTICE
Damage to the device may be caused by using the wrong cleaning agents
Do not use any abrasive, corrosive cleaning agents or strong solvents.
Ensure material compatibility.
Do not clean the product with acetone or thinner.
Never clean electrical devices with water.
Prerequisites
Safety gloves to prevent coming into contact with insulating oil
Mild detergent for cleaning the surfaces of the device
Petroleum ether or other similar mild solvent for cleaning the test chamber
Lint-free cleaning cloth