Operating Manual

PAC50H/S

Conductivity measurement

Summary of Contents for PAC50H

Page 1: ...V1 2 24 2022 EN US Operating Manual PAC50H S Conductivity measurement...

Page 2: ...10 5 1 Transport 10 5 2 Delivery inspection 10 5 3 Storage 10 6 Installation 11 6 1 Introduction 11 6 2 Mounting prerequisites 11 6 3 Installation positions 11 6 4 Installation factor 12 6 5 Mounting...

Page 3: ...ch setup 31 10 4 5 Submenu Display setup 31 10 4 6 Submenu Calibration 32 10 4 7 Submenu Identification 33 10 4 8 Submenu Diagnostics 33 11 Preventive maintenance 34 11 1 Maintenance chart 34 11 2 Cle...

Page 4: ...apply The illustrations in this manual are examples only Deviations are at the discretion of Baumer at all times 1 2 Warnings in this manual Warnings draw attention to potential personal injury or ma...

Page 5: ...erating manual n Data sheet n 3D CAD drawing n Quickstart n Dimensional drawing n Connection diagram pin assignment n IODD file n Certificates EU conformity declaration etc 1 5 Name plate The name pla...

Page 6: ...d calibration of this product may only be performed by a specialist Installation Only use the fasteners and fastener accessories intended for this product for installation Out puts not in use must not...

Page 7: ...ation Contamination may penetrate into the system and result in material damage a Always cover exposed system parts before carrying out any work e g drilling b Immediately collect any dust and chips p...

Page 8: ...a the display or using FlexProgram or the IO Link master Sensor mount is by screwing in the thread Measurements are made by the sensor s measuring hole and temperature probe Also see about this 2 Oper...

Page 9: ...out mechanical or process interfer ing components The hole in the sensor tip is surrounded by 2 coils The primary coil is supplied with AC voltage The secondary coil measures the induced current of th...

Page 10: ...can only be claimed within the applicable claims deadlines In case of externally visible transport damage proceed as follows Instruction a Do not accept the delivery or only with reservations b Note t...

Page 11: ...sensor is protected against direct sunlight n The installation site enables appropriate sensor grounding n The installation site allows for appropriate electrical sensor connection n The installation...

Page 12: ...actor depends on diameter and conductivity of the pipe socket as well as the distance from wall to sensor In case of sufficient distance to wall a 20 mm the installation factor f can be ignored f 1 00...

Page 13: ...installation in pipe systems may cause leaks Such leaks may cause a great number of hazards Therefore it is imperative to observe the following during sensor mount a Only use welding sleeves and adap...

Page 14: ...6 Installation Baumer 14 Operating Manual PAC50H S V1 Also see about this 2 Mounting prerequisites 11 6 6 Mounting example with weld in socket ZPW2 521 1 1 1 2 1 ZPW2 521 2 Leakage hole...

Page 15: ...corresponding installation parts These are marked with the logo EHEDG Certified The requirements of the 3 A Sanitary Standard are only fulfilled when using cor responding installation parts These are...

Page 16: ...ctrical connections are clean CAUTION Risk of damage to equipment by electric current NOTICE Risk of damage to equipment by contamination NOTICE Damage to seal or push in fitting Excessive tightening...

Page 17: ...Baumer Electrical installation 8 V1 PAC50H S Operating Manual 17 Also see about this 2 Pin assignment 16...

Page 18: ...ounting the sensor 13 2 Electrical connection 16 2 Parameterization using the integrated display 18 2 Parameterization via FlexProgram 22 2 Parameterization via IO Link master 23 9 2 Factory settings...

Page 19: ...navigation 28 9 3 1 Select language example The following example shows how to select the language via the sensor display Instruction a Click the additional bar on the display Button Menu appears in t...

Page 20: ...selection by clicking on Set Result The selected language has been adopted Also see about this 2 Display menu navigation 28 9 3 2 Example Change display orientation For convenient reading in any senso...

Page 21: ...color d Open submenu Display setup by pressing Enter Submenu Display setup opens e Select Orientation with the arrows Up Down Orientation is highlighted in dark color Orientation Language f Open Orie...

Page 22: ...rammer 9701 Instruction a Connect FlexProgrammer 9701 to sensor using the 5 pin M12 connector b Connect FlexProgrammer 9701 to a PC The PC must have the latest version of FlexProgrammer installed c Pr...

Page 23: ...must not be directly connected to a class B Master c Proceed with sensor parameterization Parameterization options example n Define the conductivity range n Define output damping n Define the measuri...

Page 24: ...formed in a controlled environment Instruction a Open software FlexProgram on the PC b Open setting PAC50 in the left column c Click on icon Measuring Online d Click on button Calibrate Media below th...

Page 25: ...Baumer Commissioning 9 V1 PAC50H S Operating Manual 25 INFO Helpdesk FlexProgram For more information see the Help menu FlexProgram...

Page 26: ...nductivity Concentration Details Medium Measuring point Bar chart n Additional bar Display of measuring unit analog output 4 20 mA or TAG Clicking on the additional bar opens the display menu Clicking...

Page 27: ...ect settings n Value large det n Value large n Value Temp Tag Media1 Label n Media label tag n Media label det Visual display of alarm messages ERROR App Specific 3 500 mA 149 0 C Error 0x00000006 n R...

Page 28: ...ng n Confirm selection n Confirm entries n Activation deactivation of functions Arrow buttons up down n Selection of menus and functions n Set values Arrow buttons left right n Select values n Set val...

Page 29: ...ecimal Places Define number of decimal digits assigned to the measured values Temperature unit n Temperature unit Select temperature unit Damping n Damping Select damping settings Select SEL pin n Mul...

Page 30: ...ue Value Temperature Temp at 4mA Temp at 20mA Value C F Value C F Value mA Value mA Fixed output current Value mA 0 disable Default Iout source Cond Conc Temperature Value sec Damping 4 20mA out Menu...

Page 31: ...setup is for parameterization of the sensor s the switching behavior Switch 1 n Switch 1 Selection of functions and values for the sensor switching be havior at Switch 1 Switching source Switching fu...

Page 32: ...k of display menu Pass word required 10 4 6 Submenu Calibration Calibration Restore fact calib Accept Cancel Temp offset Zero pt calibration Value C F Media calibration Resistor calibration Installati...

Page 33: ...r information Set user date n Date To indicate and enter the date 10 4 8 Submenu Diagnostics Diagnostics Statistics Min temp Max temp Uptime Total operation Demo setup Select demo mode Disabled Static...

Page 34: ...oper cleaning Unsuitable cleaning agents and methods can cause leaks and damage the rotary encoder the seals or the connections a Always check the suitability of the cleaning agent for the surface to...

Page 35: ...d of a reference sensor In this method the medium values for conductivity and temperature are measured both by an appropriate reference sensor and the sensor which is to be calibrated Deviations in th...

Page 36: ...e directly entered in the Input settings Also see about this 2 Reset user calibration 39 11 3 2 Sensor calibration with the aid of a reference medium In this method sensor calibration uses an appropri...

Page 37: ...alibration uses calibrated resistors dry calibration For doing so dein stall the sensor INFO This task can be done either using the FlexProgram or via IO Link e g BSS The following ex ample shows cali...

Page 38: ...er of measuring points calibration points per measuring range from the drop down list Calibration points per range g Insert the cable of the calibration box into the bore hole receiving the medium h C...

Page 39: ...r geometry e g by abra sion If the sensor tip changes dimensions the measured conductivity value will be different from the previous one measured with the original tip geometry Any change in geometry...

Page 40: ...11 Preventive maintenance Baumer 40 Operating Manual PAC50H S V1 Result The calibration is reset to the factory state Offsets are set to 0...

Page 41: ...r code and contact your Baumer sales company Also see about this 2 Error alarms 41 12 2 Error alarms There are two types of errors MEM ERROR App Specific 4 000 mA 24 8 C Mem error Malfunction n MEM ER...

Page 42: ...List of illustrations Baumer 42 Operating Manual PAC50H S V1 List of illustrations Ill 1 Name plate 5 Ill 2 Structure 8 Ill 3 Functional description schematic 9 Ill 4 Display 26...

Page 43: ...orientation 20 Language 19 F Functional description 9 G General information 8 I Inside cleaning 34 Installation 11 Description 18 electric 18 General 11 Installation positions 11 mechanical 18 Prerequ...

Page 44: ...www baumer com worldwide 184734091...