Caring For Your EASYpure RO
With proper care, your EASYpure RO will give
you consistently pure water. To ensure contin-
ued good service, regularly perform the follow-
ing procedures.
System Sanitization
The frequency of sanitizing depends on the rate
and amount of contamination buildup.
The membrane cartridges must be removed.
The EASYpure RO can easily be completely
sanitized by utilizing the sanitization cartridge
(Part Number D50245). To accomplish this, fol-
low these step-by-step instructions:
1.
Turn off the EASYpure RO. Turn off an
EASYpure unit or NANOpure fed by the
EASYpure RO. Depressurize the
EASYpure unit or NANOpure by open-
ing the drawoff valve (auxiliary drawoff
on NANOpure) and allowing water to
flow until no more exits the unit.
Disconnect the EASYpure unit or
NANOpure from the power supply.
2.
Close the inlet valve on the EASYpure
unit. Disconnect from the EASYpure RO
the tubing attaching the EASYpure RO
to the EASYpure unit. Disconnect this
same tubing from the EASYpure unit.
Direct one end of the tubing to drain.
Reattach the other end to the EASYpure
RO. Allow the tank to drain.
3.
If you are feeding a NANOpure, remove
from the NANOpure the tubing from the
"T" on the accessory 30 Liter Reservoir.
Drain the accessory 30 Liter Reservoir
by opening the spigot.
22
Maintenance
Warning
Disconnect from the power supply
prior to maintenance and servicing.
Refer servicing to qualified personnel.
Ensure all piping connections are tight
to avoid leakage of chemicals.
Ensure adequate ventilation when
using chemicals for
cleaning.
Follow carefully the manufacturer’s
safety instructions on labels of chemi-
cal containers and Material Safety
Data Sheet (M.S.D.S.)