6. Installation of Projection System
OverView D 50" – DOC-3303-5 – installation manual – Revision 10 – September 2006
September 2006
September 2006
September 2006 ________________________________
________________________________
________________________________
________________________________________________________
________________________
________________________
________________________6-11
6.4
Cabling
The following figure shows a block diagram about the cabling.
pu
iu
pu = projection unit
iu = illumination unit
data
data
power
power
main
power
external
fan
dvi
external
IR - sensor
Figure 6-10
Block diagram
Figure 6-11
Overview cabling
Pos
Description
A
Projection unit
B
Illumination unit
1
DVI connector
2
External infrared sensor
3
COM in
4
COM out
5
Data
6
Power
Table 8
When using optical(fibre) data cables, please note that they may not be bent to a
curvature with a bend radius less than 15.4cm!
As soon as the illumination unit and the projection unit are connected by the signal cable
and/or the pow er cable, DO NOT disconnect either of these cables under operation or in
standby!
Disconnect only when the system is completely powered off (all LEDs are off)!
Otherwise the trinamic board of the IU will be damaged.