![Barco Nio 2MP LED User Manual Download Page 24](http://html1.mh-extra.com/html/barco/nio-2mp-led/nio-2mp-led_user-manual_4109155024.webp)
K5903058 /11
Nio 2MP LED Display
24
4. Select
Enabled/Disabled
as desired and confirm.
To specify the hibernate time-out:
1. Bring up the OSD main menu.
2. Navigate to the
Configuration > Power Management
menu.
3. Enter the
Hibernate Timeout
submenu.
4. Set the time-out value as desired and confirm.
4.8 Luminance target
About the luminance target
The luminance target of your Nio 2MP LED Display is adjustable over a predefined range. When you change
the luminance target, the display will adjust its backlight to reach the target.
To set the luminance target:
1. Bring up the OSD main menu.
2. Navigate to the
Configuration > Calibration
menu.
3. Enter the
Luminance Target
submenu.
4. Set a luminance target value as desired and confirm.
The default, factory DICOM calibrated luminance value is available in the technical specifications
table. The guaranteed backlight lifetime is valid for this setting.
4.9 Viewing modes
About viewing modes
The Nio 2MP LED Display can be used in two viewing modes:
•
Diagnostic:
This mode provides the full calibrated luminance and is intended for using the display for
diagnostic purposes.
•
Text:
In this mode, the luminance is reduced to approximately half of the luminance. This is intended for
using the display with office applications such as word processing.
Please note that text mode is not persistent, once powered off, the unit will restart in diagnostic mode.
The diagnostic mode should always be selected when the Nio 2MP LED Display is intended to be
used in a diagnostic environment.
To select a viewing mode:
1. Bring up the OSD main menu.
2. Navigate to the
Configuration > Calibration
menu.
3. Enter the
Viewing Mode
submenu.
4. Select
Diagnostic/Text
as desired and confirm.
4.10 Display functions
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
Summary of Contents for Nio 2MP LED
Page 1: ...ENABLING BRIGHT OUTCOMES MDNC 2221 User Guide Nio 2MP LED Display...
Page 5: ...5 K5903058 11 Nio 2MP LED Display Welcome 1...
Page 8: ...K5903058 11 Nio 2MP LED Display 8 Welcome...
Page 9: ...9 K5903058 11 Nio 2MP LED Display Installation 2...
Page 16: ...K5903058 11 Nio 2MP LED Display 16 Installation...
Page 17: ...17 K5903058 11 Nio 2MP LED Display Daily operation 3...
Page 20: ...K5903058 11 Nio 2MP LED Display 20 Enter Cancel Standby IEC 60417 5009 Daily operation...
Page 21: ...21 K5903058 11 Nio 2MP LED Display Advanced operation 4...
Page 28: ...K5903058 11 Nio 2MP LED Display 28 Advanced operation...
Page 29: ...29 K5903058 11 Nio 2MP LED Display Maintenance 5...
Page 32: ...K5903058 11 Nio 2MP LED Display 32 Maintenance...
Page 33: ...33 K5903058 11 Nio 2MP LED Display Important information 6...
Page 50: ...K5903058 11 Nio 2MP LED Display 50 Important information...
Page 51: ...51 K5903058 11 Nio 2MP LED Display Troubleshooting 7...
Page 53: ......
Page 54: ...www barco com K5903058 11 2021 12 20 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...