33
R5913216 /03
MDPC-8127
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown: Time
elapsed since latest calibration, Display Function, Ambient Light Compensation and Reading Room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Status Report
to make the information visible on the screen.
4.16.3 DICOM compliance check
About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Compliance Check
to start the compliance check.
Warning:
Pressing a key during the compliance check will abort the check.
4.16.4 DICOM calibration
About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
To start DICOM calibration:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM calibration
to start the calibration.
Warning:
Pressing a key during calibration will abort the calibration, previous values will be
restored.
Note:
After calibration, the compliance check will start automatically.
4.16.5 Reset DICOM calibration
About reset DICOM calibration
It is possible to restore the original (factory default) DICOM curve.
Summary of Contents for MDPC-8127
Page 5: ...5 R5913216 03 MDPC 8127 Welcome 1...
Page 12: ...R5913216 03 MDPC 8127 12 Welcome...
Page 20: ...R5913216 03 MDPC 8127 20 Installation and setup...
Page 21: ...21 R5913216 03 MDPC 8127 Daily operation 3...
Page 24: ...R5913216 03 MDPC 8127 24 Daily operation...
Page 37: ...37 R5913216 03 MDPC 8127 Cleaning your display 5...
Page 39: ...39 R5913216 03 MDPC 8127 Important information 6...
Page 51: ......
Page 52: ...www barco com R5913216 03 2021 09 10 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...