3. Operation
3.6
Luminance adjustment
To adjust the luminance
1. Bring up the OSD main menu.
2. Navigate to the
Adjustments
>
Luminance
menu.
3. Set a luminance value as desired and con
fi
rm.
3.7
sRGB color space
About sRGB color space
The sRGB color space combines a display function and white point selection and is designed to match
typical home and of
fi
ce viewing conditions. It is widely used in most computer applications.
When selecting
sRGB
, the
Display function
and
White point
selection options in the
Ad-
justments
menu will be disabled.
To select sRGB color space
1. Bring up the OSD main menu.
2. Navigate to the
Adjustments
menu.
3. Select
sRGB
and con
fi
rm.
3.8
Display functions
Display function selection is disabled when
sRGB
is selected in the
Adjustments
menu.
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been de
fi
ned. These functions emphasize on these parts containing crucial diag-
nostic information by correcting the native panel behavior.
The available display functions for your Nio Color 2MP 23" are:
•
Native
: If you select Native, the native panel behavior will not be corrected.
•
DICOM
: DICOM (Digital Imaging and Communications in Medicine) is an international standard that
was developed to improve the quality and communication of digital images in radiology. In short, the
DICOM display function results in more visible grayscales in the images. Barco recommends selecting
the DICOM display function for most medical viewing applications.
The DICOM display function applies ambient light compensation (ALC) taking the ambient light con-
ditions of your reading room into account. The available reading room options are:
-
Darkroom
: Selects DICOM calibrated function, optimized for darkroom conditions (0 Lux)
-
Of
fi
ce
: Selects DICOM function optimized for of
fi
ce conditions (60-180 Lux)
-
Operation Room
: Selects DICOM function optimized for operating room conditions (300-400 Lux)
K5902134 NIO COLOR 2MP 23" 05/07/2017
15