5.4 EMC notice
General information
This device is for use in professional healthcare facility environments only.
With the installation of the device, use only the delivered external cables and power supply or a spare part
provided by the legal manufacturer. Using another can result in a decrease of the immunity level of the device.
WARNING:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING:
Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the MDMX-
MNA, including cables specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
CAUTION:
This equipment is not intended for use in residential environments and may not provide
adequate protection to radio reception in such environments.
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B
is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or re-
orienting the equipment.
Electromagnetic emissions
The MDMX-MNA is intended for use in the electromagnetic environment specified below. The customer or the
user of the MDMX-MNA should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
Guidance
RF emissions
CISPR 11
Group 1
The MDMX-MNA uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
Class A equipment is equipment
suitable for use in all locations other
than those allocated in residential
environments and those directly
connected to a low voltage power
supply network which supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
The MDMX-MNA belongs to the medical graded Group 1 equipment: this means it makes no use of radio-
frequency energy in the frequency range 9 kHz to 400 GHz that is intentionally generated and used or only
used locally, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of
material, for inspection/analysis purposes, or for transfer of electromagnetic energy.
The MDMX-MNA complies with appropriate medical EMC standards on emissions to, and interference from
surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Summary of Contents for MDMX-MNA
Page 4: ...R5913897 00 MDMX MNA 4...
Page 5: ...5 R5913897 00 MDMX MNA Welcome 1...
Page 7: ...7 R5913897 00 MDMX MNA Image 1 1 MDMX 22400 Option GNTB Welcome...
Page 10: ...R5913897 00 MDMX MNA 10 Welcome...
Page 11: ...11 R5913897 00 MDMX MNA Parts controls and connectors 2...
Page 14: ...R5913897 00 MDMX MNA 14 Image 2 4 Parts controls and connectors...
Page 15: ...15 R5913897 00 MDMX MNA Daily operation 3...
Page 19: ...19 R5913897 00 MDMX MNA Cleaning and disinfection 4...
Page 21: ...21 R5913897 00 MDMX MNA Important information 5...
Page 38: ...R5913897 00 MDMX MNA 38 Important information...
Page 39: ......