Turbidity measuring

 TURBICUBE

 

BAMO  IER GmbH 

 Pirnaer Straße 24 

 D-68309 Mannheim 

Tel. +49 (0)621 84224-0 

 Fax: +49 (0)621 84224-90 

 www.bamo.de 

 [email protected]  

SU3332a.doc   01/13 

1

 

 

 

 

 

 

Installation, initial start-up and maintenance may only be performed by trained personnel! All applicable 
European and national regulations regarding installation of electrical equipment must be adhered to. 

 

The device may only be connected to supply power which complies with the specifications included in the 
technical data and on the serial plate! 

 

The device must be disconnected from all sources of power during installation and maintenance work! 

 

The device may only be operated under the conditions specified in the operating instructions! 

 

Functions Description: 

 

The turbidity measuring device determines the turbidity of liquids using the light absorption and scattered light 
measuring method. 
The inline measuring cell at the 

sensor set

 makes use of two IR transmitters and one IR receiver in pulsating light 

mode (IR measuring method according ISO 7027). 
The interconnected BAMOPHOX TUR measuring amplifier analyses the 4 to 20 mA signal generated by the sensor 
set. 

 

Installing the Sensor Fixture: 

 

Observe: 
The fixture must be installed vertically! 
Best installation is the mounting in a ascending tube with slow down section 600 mm in front and 400 
mm behind the turbidity measuring device 

 

The sensor fixture must be completely filled with liquid medium during the performance of measurements. 
Install to the vertical portion of a siphon trap if necessary. 

 

Air bubbles distort measurement results. 

 

Glasses must be clean – clean at regular intervals as required. 

Safety Precautions: 

Summary of Contents for IER Turbicube 20

Page 1: ...Functions Description The turbidity measuring device determines the turbidity of liquids using the light absorption and scattered light measuring method The inline measuring cell at the sensor set mak...

Page 2: ...mark if the measuring circuit is also connected to the auxiliary power supply of the load will increase to additional 24 VDC x 20mA 0 48W Terminal housing PBT IP 65 Ambient temperature 5 45 C Medium t...

Page 3: ...OFF OFF OFF OFF OFF DIP switch 5 and 6 without function Normal position OFF armatures with nominal diameter DN65 DN100 Troubleshooting Measuring current 22 mA overflow Measuring current 0 mA wrong DI...

Page 4: ...Module PCB Elektrical Connection Observe The measuring output 4 20 mA is a current drain that means the output channel has the same electrical behavior as a variable resistor A active power supply fo...

Page 5: ...e fixture by unscrewing the sleeve nuts screw the ETR glasses with the included tool and remove clean the lens do not use abrasive cleansers which may cause scratching tighten the ETR glasses with the...

Page 6: ...ng is required as part of the respective quality assurance system for test equipment monitoring calibration can be checked with the calibration rods and the device can be readjusted if necessary A sui...

Page 7: ...bration rod from the box Insert the calibration rod into the fixture Mount the S2 transmitter to the calibration rod observe the locking pin Insert the calibration rod with attached transmitter comple...

Page 8: ...SW1 setpoint on the plate in the box of the KN D calibration rod Actual value measured value Setpoint actual value deviation Less than 5 measuring instrument is OK Greater than 5 adjust actual value...

Page 9: ...onnection both side flange DIN 2501 DN20 d25 240 240 300 300 DN25 d32 240 240 300 300 DN32 d40 240 240 300 300 DN40 d50 240 240 300 300 DN50 d63 262 262 342 300 DN65 d75 314 314 342 300 DN80 d90 347 3...