![Balluff BTL7-A5-M Series User Manual Download Page 38](http://html1.mh-extra.com/html/balluff/btl7-a5-m-series/btl7-a5-m-series_user-manual_3830633038.webp)
english
12
6
Calibration procedure
Output
gradient
BTL
Unit
Min.
value
Null
value
Indication for
adjustment
Indication
for teach-in
End
value
Max.
value
Error
value
Rising
BTL7-A…
V
–0.5
0
2.0
4.0
+10.0
+10.5
+10.5
BTL7-C…
mA
0
0
6.0
12.0
20.0
20.4
20.4
BTL7-E…
mA
3.6
4.0
6.0
12.0
20.0
20.4
3.6
Falling
BTL7-A…
V
+10.5
+10.0
2.0
4.0
0
–0.5
–0.5
BTL7-C…
mA
20.4
20.0
6.0
12.0
0
0
20.4
BTL7-E…
mA
20.4
20.0
6.0
12.0
4.0
3.6
3.6
Tab. 6-1: Value table for teach-in and inverting
6.1
Programming inputs
Programming inputs La and Lb must be used in order to
make settings. A programming input at 10 to 30 V
corresponds to activation (high active).
The Balluff calibration box BTL7-A-CB02-… can be used
for this (see Accessories on page 20).
Automatic deactivation!
If no signals are transmitted via the
programming inputs for approx. 10 min,
programming mode is automatically ended.
6.2
Calibration procedure notes
Prerequisites
– Programming inputs are connected.
– The BTL connected to the system controller.
– Voltage or current values from the BTL can be read
(using a multimeter, the system control or the
calibration box).
8-pin
YE
GY
PK
RD
GN
BU
BN
WH
YE
GY
PK
RD
GN
BU
BN
WH
BTL
Calibration box
Supply
Adapter cable with
connector (luster terminal)
6-pin
8-pin luster strip
Fig. 6-1: Connecting the calibration box BTL7-A-CB02-S…
Values for null and end point
– Any desired position of the float can be used as the null
or end point. However, the null and end points may not
be reversed.
– The absolute null and end points must lie within the
minimum or maximum range of what can be output
(see value table).
The last set values are always saved, regardless
of whether the setting was ended using the
programming inputs or automatically after
10 min. have expired.
Value table for teach-in and inverting
The following examples refer to BTLs with
0…10 V or 4…20 mA output.
BTL7-A/C/E5 _ _ -M _ _ _ _ -SF-F _ _
Magnetostrictive Linear Position Sensor – Rod Style
Summary of Contents for BTL7-A5-M Series
Page 2: ...www balluff com...
Page 3: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Betriebsanleitung deutsch...
Page 4: ...www balluff com...
Page 26: ......
Page 27: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ User s Guide english...
Page 28: ...www balluff com...
Page 50: ......
Page 51: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Notice d utilisation fran ais...
Page 52: ...www balluff com...
Page 74: ......
Page 75: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manuale d uso italiano...
Page 76: ...www balluff com...
Page 98: ......
Page 99: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Manual de instrucciones espa ol...
Page 100: ...www balluff com...
Page 122: ......
Page 123: ...BTL7 A C E5_ _ M_ _ _ _ SF F_ _ Instrukcja obs ugi polski...
Page 124: ...www balluff com...