Chapter 4. Error Codes and Troubleshooting
44
PanaFlow™ LC User’s Manual
[no content intended for this page]
Peak Index Down
Threshold peak of the
downstream transmit
correlation signal
•
For pipe sizes greater
than 1 inch, index should
be between 400 - 700.
•
For pipe sizes less than 1
inch, the index should be
between 150 - 350.
•
For pipe sizes greater than 1 inch, if
the index <400 or >700 then there
is an indication of problem with
receive window location.
•
For pipe sizes less than 1 inch, if the
index <150 or >350 then there is an
indication of problem with receive
window location.
Wall Time
Transit time inside the
pipe wall
N.A
If the value is negative, then there is
an indication of problem with the
configuration parameters.
Lining Time
Transit time inside the
pipe lining
N.A
If the value is negative, then there is
an indication of problem with the
configuration parameters
Signal Quality Up
Signal quality of the
upstream transducer
>1000
<1000
Signal Quality
Down
Signal quality of the
downstream
transducer
>1000
<1000
Amplitude Up
Signal amplitude of the
upstream transducer
>14 and <32
>32 or <14
Amplitude Down
Signal amplitude of the
downstream
transducer
>14 and <32
>32 or <14
Table 8: Diagnostic Parameter Description and Health Indicators
Parameter
Description
Good
Bad
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...panametrics com BH029C11 EN C PanaFlow LC User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 64: ...Appendix D CE Mark Compliance 56 PanaFlow LC User s Manual no content intended for this page...
Page 66: ...Warranty 58 PanaFlow LZ User s Manual no content intended for this page...
Page 67: ......